SJM Confirm DM2100 DM2102 (ICM)

Conditional

ST. JUDE MEDICAL, INC

Implantable cardiac monitor

  • Testing has demonstrated that the St. Jude Medical™MR Conditional SJM Confirm™ ICM is conditionally safe for use in the MRI environment when used according to the instructions in this manual.

  • MR Conditional Contraindications: Patients with abandoned cardiac hardware including leads, lead extenders, or lead adaptors are contraindicated for an MRI scan.

  • Patients with non-MR Conditional implantable devices are contraindicated for an MRI scan.

  • Patients with a St. Jude Medical™ SJM Confirm™ ICM that has been implanted for less than 6 weeks are contraindicated for an MRI scan. The 6 week post-implant waiting period allows time for the implant pocket and wound to heal which minimizes the effects of "tugging" on the device caused by magnetic fields.

  • Patients with this device must not be placed on their side within the MRI bore, this position is contraindicated