No other components of Freedom SCS System (e.g. Transmitter Assembly,battery charger, needles, stylets, guidewire, trial lead) may be taken into the MR system room.
Remove the Transmitter Assembly from the patient before entering the MR system room. The strong magnetic field of the MR system could attract or otherwise damage the unit and may cause serious harm or damage to the Transmitter Assembly and/or the MR system.
The patient should be conscious during the MRI procedure. Monitor the patient both visually and audibly. Check the patient between each MR imaging sequence. Discontinue the MRI examination immediately if the patient is unable to respond to questions or reports any problem.